Activity
From 12/28/2014 to 01/26/2015
01/19/2015
- 03:38 PM Support: RE: top and bottom half sensor temp
- Ok great thanks! Just wanted to double check.
- 03:16 PM Support: RE: top and bottom half sensor temp
- Hi Kyle,
This is why we use an external thermistor attached to the sensor for temperature reporting. We have obse... - I have noticed that there is a big temp difference between the top and bottom half of the sensor.
Example:
Vpt...
01/08/2015
- 10:41 AM Support: RE: High/Low gain mode using 8bpp-12bpp
- Mike, is there a similar bias in the 8bit and 12bit modes?
If so what is it?
Does 12bit pack into 11bits in those ...
12/30/2014
- 01:11 PM Support: RE: High/Low gain mode issue using 16bpp
- Ah, ok well that explains it! No problem, just good to know what is happening.
Thanks.
Kyle - 12:50 PM Support: RE: High/Low gain mode issue using 16bpp
- There is a 20 count bias in the data. It's a bug in the mode you are using.
During calibrated high/low gain blend... - 12:38 PM Support: RE: High/Low gain mode issue using 16bpp
- I use MATLAB to look at and process the data. Knowing that the data is packed into the lower 11 helps answers my ques...
12/29/2014
- 08:04 AM Support: RE: High/Low gain mode issue using 16bpp
- Hi Kyle,
Are you using XCAP (the epix frame grabber) or MityViewer to look at the data?
When you switch to sing...
Also available in: Atom
Go to top