MityCAM-120MXS Firmware¶
This page documents the firmware revisions of the MityCAM-120MXS camera firmware. Currently, the same firmware is loaded on all cameras with the 120MXS sensor, regardless of capability (Mono, Color, RGBIR).
In general, Critical Link does not recommend upgrading firmware unless there is a specific feature that has been added or performance issue resolved that you require. Failure to install compatible firmware or to properly follow the installation procedure may leave the camera in an unusable state and may possibly damage the unit. If you unsure or have any questions about the upgrade, please contact your Critical Link representative prior to starting the upgrade procedure.
MityCAM-120MXS Firmware Update instructions:
- Copy the update
.run
file to your local filesystem. - Connect the USB 2.0 cable https://media.digikey.com/photos/Qualtek%20Photos/3021009-06.JPG to the camera, connect it to your PC and power on the camera. Wait about 60 seconds for the camera to boot.
- Open a web browser and enter http://10.1.47.2/index.py in the address bar
- You should see a "MityCAM Maintenance Page" note the value listed for "Build Info" under the Camera Information section.
- At the bottom of the page, select "Browse" or "Choose File" in the Firmware Upgrade section and select the correct
.run
file. - Click "Upload"
- The status screen displays the progress of the upgrade. Note that you may see "Upgrade Failed" messages for components that do not need to be upgraded.
- Please cycle on the camera after the upgrade is complete.
- After the camera has rebooted, return to http://10.1.47.2/index.py and verify that the "Build Info" has changed.
Release | Date | .run file |
---|---|---|
52 | 2020-03-13 | update_camera_20200313_155552.run |
47 | 2020-02-24 | update_camera_20200218_104314.run |
52 [2020-03-13]¶
Change Log¶
The following toolchain was used to build the firmware.
- FPGA built with Quartus Prime PRO 18.1
- ARM software built with gcc 7.3.0 ARM toolchain
Added Features¶
- None
Bugs Fixed¶
- Frame sync was being lost in some cases when the image was in saturation. This would look like sporadic bad columns of data or a vertical rolling image.
Known Issues¶
- None
47 [2020-02-24]¶
Change Log¶
Please see updated user manuals for information about newly added features.
The following toolchain was used to build the firmware.
- FPGA built with Quartus Prime PRO 18.1
- ARM software built with gcc 7.3.0 ARM toolchain
Added Features¶
- None
Bugs Fixed¶
- Fixed saturation issue with camera image which resulted in strange colors in the camera image in some lighting conditions (Mono, Color, RGBIR)
Known Issues¶
- If using the IR Subtraction feature in the GenTL viewer, please update to the newest GenTL Viewer (TBD) after installing this firmware upgrade.